How supplied: The enclosed vial(s) contains a concentrated solution of beads suspended in distilled water. If red beads are being used for retrograde tracing of neuronal pathways, the solution can be used as is, or diluted. In rat visual cortex, dilutions of 1:4 do not appear to reduce the quality or extent of retrograde labeling when using red beads. However, for initial experiments we strongly recommend using the solution full strength. In addition to distilled water, standard salt solutions (NaCl, KCl) can be used as diluents. The green beads, as supplied, are completely prepared for retrograde tracing experiments. Dilution of green beads is not recommended.
Storage: The bead solution should be stored in a humidified container, in a refrigerator, to prevent evaporation. Do not freeze! Beads that have been frozen will not work, and cannot be rescued. If the beads dry out, they cannot be reconstituted. No shelf life has been established for this material, but, when properly stored, it remains good for several years.
Application: Beads are best injected using pressure (e.g. a 1 ml
Survival times: The minimum effective post-injection survival time in most warm-blooded vertebrate systems is approximately 24 hours. Labeling intensity increases with longer survival, up to 48 hours. After 48 hours, no increase (or decrease) in labeling intensity is observed. These values may be considerably different in cold-blooded animals, and initial survival times of a week are recommended. The maximum survival time has not been established, but labeling is unchanged in either quality or extent even after 14 month survival times. Cells probably are permanently marked. No toxic effects on either animals or neurons have been observed.
Fixation and processing: Standard fixation is a 0.1 M phosphate buffer wash followed by 4% paraformaldehyde in 0.1 M phosphate buffer (pH 7.4). Glutaraldehyde will produce substantial tissue autofluorescence which may obscure bead-labeled neurons, and should be avoided if possible. Green beads will be completely invisible in glutaraldehyde fixed material. Frozen sections are collected in phosphate buffer, mounted on gelatin-coated slides, and air-dried. After complete drying, slides can be cleared in xylene for 1 minute, and covers lipped with Fluoromount or Krystalon. Fluoromount is available from Atomergic Chemetals Corp.,
Observation: The dye in the red beads is rhodamine, thus any fluorescence filter set for rhodamine can be used. Some older Nikon rhodamine filter sets give a very high background, which can make labeled cells invisible. Good results have been obtained with both Zeiss and Leitz standard rhodamine filters. For green beads, a wide-band fluorescein filter works well. Filter sets for Lucifer yellow give a more intense signal, but at the expense of higher background. Narrow-band fluorescein filters will give a much weaker signal than a broad-band filter. Beads do not fade appreciably even after long periods of observation or photomicrography.
Labeling is usually not visible with low power, low numerical aperture dry objectives (e.g. X4, X10). If cells are strongly labeled, a X10 immersion objective (numerical aperture of 0.4 or greater), or higher power dry objectives, will usually reveal the cells. However, frequently a X25 immersion objective will reveal very clearly labeled cells that lower power objectives miss. These caveats are especially true for green beads. Before deciding that an experiment did not work, examine sections in the vicinity of the injection site with immersion objectives. Numerous locally labeled cells should be present.
Additional information for green bead users: In work that has been done so far, it appears that younger animals transport the label better than older animals. In addition, tissue autofluorescence is lower in the younger animals. Therefore, it is advisable to use younger animals, if possible, in experiments involving green beads.
Because the green beads are excited at shorter wavelengths than red beads, tissue autofluorescence is a greater problem. Therefore, efforts to minimize autofluorescence will produce a better contrast signal. Ways to reduce autofluorescence include: 1) using thinner sections (e.g. 30 um rather than 40 or 50) 2) using younger animals, and 3) examining sections promptly after coverslipping (background increases over time).